FAQ

   
Q. How long is Delsaro Scientific Solutions pipette performance warranty following service?

A. In comparison to other unauthorised service providers, we guarantee that parts are replaced with new, genuine spares as recommended by the manufacturer. This commitment to quality allows us to warranty our service for 12 months

Q. How often should I return my pipettes for service & calibration?

A. The frequency of both service and re-calibration is based on risk assessment. The risk assessment is likely to consider the frequency of the use of the pipette, the corrosiveness of the liquid being pipetted and the risk to the business if the pipette goes out of calibration. This would normally result in the decisions on how often being set between 3 months and a year.

Q. What can I do in my laboratory to maintain my own pipettes?

A. As a population, pipettes checked annually without any preventative maintenance or regular servicing procedures in place, will have a 20% or more failure rate.
Over 60% of pipette failure is attributable to sealing system defects. This is predictable but preventable. Under normal use and based on manufacturer’s recommendations seals should be replaced at least once per year, while tip holders and pistons should be replaced every three to five years. Seals, o-rings, tip holders, tip ejectors and the volume adjustment knobs can easily be changed in your own laboratory without affecting the calibration of your pipettes There are however a number of parts that when changed can alter the pipettes calibration. These parts include the piston and piston assembly, friction o-ring, operating rod and volumeter. When these need changing it is best to return your pipettes to your service provider for replacement and pipette re-calibration.
Prevention is better than cure. It is recommended that you check your pipettes daily before use and establish a routine for basic servicing. There are a number of basic care checks for your pipettes that can be performed in your own laboratory to maintain correct functionality in between full servicing. These include a simple functionality check and a visual leak test.

Q. What is the Gravimetric Method?

A. The Gravimetric test is the most commonly used method for pipette calibration by pipette manufacturers and international standard organisations.
It is based on the determination of the weight of water samples delivered by the pipette. Implementation of this method requires the strict monitoring of environmental conditions and the systematic use of adequate and controlled equipment. Ambient temperature, barometric pressure and relative humidity must all be checked and regulated.
Conversion of weight to volume must take into account the density of the liquid as well as evaporation during the cycle time. For each measurement, the corresponding volume (Vi) can be calculated as follows: Vi = (Wi + e) Z.
Wi is the weight as read on the balance; e is the mean evaporation loss during the cycle time, Z expressed in µl/mg, is a conversion factor incorporating density of water buoyed in air, at test temperature and barometric pressure.
Always calibrate to ISO standards.

Q. Why do you change the seal and O-ring on every service & calibration?

A. We always replace seals & O-rings during service as recommended by Gilson to maintain PIPETMAN’s high levels of accuracy & precision.
Other service agencies who do not use original Gilson spare parts, or who wish to reduce costs, may use grease to obtain an air-tight seal rather than replacing parts. PIPETMAN is not designed as a greased system and this quick fix will have long term detrimental effects on pistons and tip holders with a loss of calibration over time.

Q. Why is it important to check pipette linearity?

A. For mechanical reasons, pipettes can only deliver samples accurately over a limited range of volumes for each particular model.
Ideally, the volume of liquid delivered in that range follows a precise linear relationship to that selected. It is essential to check this linearity when the pipette is serviced to ensure correct functionality at both the upper and lower limits of the pipettes capacity.
Even with Delsaro Scientific Solutions’ most basic level of service, volumes are checked at both extremes of a pipettes range. Our standard service level is 2.2; meaning two volumes are checked with two weightings’ to ensure both accuracy and linearity.
To increase their throughput some other service agents only perform checks at a single mid-point volume with three or five weighing, and even issue calibration certificates based on these minimal readings.
As the majority of pipettes are variable volume how can you ensure that a pipette is linear across its specification range by testing only one volume?
Always ask what level of service you are getting for your money.

Q. Why should Gilson PIPETMAN NOT be greased?

A. If you apply grease to the internal parts of a pipette, it will need to be removed with a solvent for cleaning during servicing. The amount of grease re-applied may vary and could alter the calibration of the pipette.
PIPETMAN maintained in the correct way, as specified by Gilson, using original manufacturer’s spare parts, should never require grease to be applied. PIPETMAN can be cleaned and seals and O-rings replaced without altering the calibration.
Delsaro Scientific Solutions always replaces these critical dry-seal parts during servicing to maintain PIPETMAN’s high levels of accuracy and precision. Other service agencies who do not use original Gilson spare parts or who wish to reduce costs, rather than replacing parts, may use grease to obtain an air-tight seal. Grease may harbour contaminants within the pipette, introduced through aerosols or when pipetting volatile liquids. These are retained in the grease and may advance corrosion of internal pipette parts such as pistons.
Such damage may increase costs as additional parts are required more frequently than if the pipette is correctly maintained. PIPETMAN is not designed as a greased system and this quick fix will have a long term detrimental effect on the pipette with a loss of calibration over time.

Don’t let anyone grease your PIPETMAN!

Q. Why should I ensure that genuine manufacturer’s spares are used by my service agent?

A. The pipettes manufacturers’ warranty and specification can only be guaranteed when genuine manufacturers’ replacement spare parts are used with service and calibration performed by an authorised service organisation.
Delsaro Scientific Solutions is one of only a few Gilson approved pipette service organization in the UK and Ireland, and as such is the only one to guarantee use of genuine manufacturers spare parts.
Gilson manufactured spares are subject to the highest levels of quality control and are checked and warranted. They ensure the use of quality materials such as PVDF for the best chemical resistance, compatibility and durability. The perfect size dimensions of pistons, tip-holders and tip ejectors guarantee maximum accuracy and precision plus trouble free ejection of tips.
Now you can insist on genuine spare parts for your pipette maintenance as all Gilson pipette tip-holders, pistons and tip ejector heads are embossed with the Gilson Trident logo, guaranteeing their authenticity.

Q. How old is my pipette?

A. All Gilson pipettes carry a serial number which identifies the pipette and the date of manufacture (see example below).

Example of a serial number A10369H

The first letter represents the year of manufacture.
The number is the individual production number.
The last letter corresponds to the month of manufacture.

 

Letter

A
B
C
D
E
G
H
J
K
L
M
N
P
Q
R
S
T
U
W
X
Y
Z

Year

1984
1985
1986
1987
1988
1989
1990
1991
1992
1993
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005

Letter

A
B
C
D
E
G
H
J
K
L
M
N

Month

January
February
March
April
May
June
July
August
September
October
November
December

 
Q. What are the main causes of leaks?A. There are various reasons for leaks, the most common of which are;

Tip holder is scratched or damaged
Use of non-Gilson tips
Use of non-Gilson seals
Vapour pressure from organic solvents

Q. How do organic solvents cause leaks?

A. When an organic solvent is used with an air-displacement pipette, leaks may occur. These leaks are caused by the difference between the vapour pressure of the solvent and pressure of the air cushion between the piston and the sample.
Solution:

  • Use a positive-displacement pipette (Microman), because it has no air cushion.
  • If you are using an air-displacement pipette, saturate the air cushion of your pipette with solvent vapour by aspirating and distributing solvent repeatedly. The leak will stop when pressure equilibrium is reached.
Q. How can I prevent piston corrosion?

A. After contact with corrosive liquids, a piston should be cleaned with alcohol and a soft tissue. Take care to avoid shocks or scratches.

Q. Where do I send my pipettes for repair?

A. Send your pipettes for repair to:

Delsaro Scientific Solutions Ltd
Bedford i-Lab
Stannard Way
Priory Business Park
Bedford
MK44 3RZ

Remember to enclose the Pipette Service Form with your pipettes when shipping.
Please note: To ensure the safety of all personnel, pipettes MUST be free of biological and radioactive hazards.

Q. Why do I need a pipette holder?

A. To prevent corrosion, contamination or breakage. Always store your pipettes vertically to prevent liquids from running inside the shaft of the pipette. Do not leave your pipette lying on the workbench where it can come into contact with chemicals or fall off and break.

Q. How do I prevent aerosol contamination?

A. Use a Pipetman pipette with Diamond sterilized filter tips when faced with at least two of the following situations at the same time:

- working in sterile conditions
- pipetting aqueous samples
- avoiding cross-contamination

Use a Microman pipette with sterilized capillaries and pistons when faced with at least two of the following situations at the same time:

- working in sterile conditions
- pipetting viscous samples
- avoiding cross-contamination

Q. What are the best replacement parts for my Gilson Pipettes?

A. Only genuine Gilson replacement parts should be used for your Gilson pipettes.

Q. What are published specifications?

A. Specifications are established by the manufacturer. They guarantee, in terms of accuracy and precision, the performance of all pipettes of a given brand and a given model at a certain volume setting.

Accuracy - is the ability of a measuring instrument to give responses close to a true value.
Precisionis the ability of an instrument to provide closely similar responses (measurements). Precision is also referred to as repeatability and/or reproducibility.

Q. The tips do not fit well, what should I do?

A. Always use Gilson calibrated pipette tips for Gilson pipettes.

Push upward on the tip-ejector to make sure it is positioned properly.
Clean the tip holder with alcohol. If it is worn or has been chemically attacked, order a new part.